Split Medium temperature surface source blackbody furnace
The surface source blackbody radiation source/blackbody calibration furnace/blackbody infrared calibrator is the working standard for infrared radiation. It is
- Enquiry-based quotation
- Technical & calibration support
- Worldwide enquiries welcome
Description
Product Introduction
The surface source blackbody radiation source/blackbody calibration furnace/blackbody infrared calibrator is the working standard for infrared radiation. It is widely used in the process testing and qualification testing of technical indicators in the development process of infrared detectors. It can achieve the uniformity of the photosensitive surface response of the infrared focal plane device. At the same time, the surface source blackbody radiation source, as the core equipment of the infrared imaging collimation device, cooperates with the infrared collimation optical system and infrared targets to complete the testing and performance evaluation of key technical indicators such as the infrared imaging system. It can meet the testing needs of infrared imagers in infrared imaging guided weapons developed in our country, and can meet the testing needs of infrared thermal imaging cameras widely used in medical, construction, security monitoring, environmental detection and other industries.
Product Feature
- More portable and lighter. Higher accuracy.
- Human-machine interface, Chinese display, touch operation, beautiful and elegant.
- One-click setting of temperature, easy to operate. Rapid heating and temperature control.
- According to customer requirements, the language interfaces of multiple countries and regions can be copied to facilitate export.
- Temperature setting, simple operation.Rapid heating, temperature control.
- Product advantages
- The appearance is novel and the furnace and temperature controller are separated.
- The heating element is insulated from the shell, and special software is used for pressure-limited heating, so no high-power power transformer is required.
- Automatic temperature rise and temperature control is adopted, which is safe and reliable, with fast heating speed and good temperature stability.
- It is easy to use and operate, small in size and light in weight, easy to carry, and is not only suitable for laboratory calibration, but also for on-site calibration.
- Large surface source isothermal blackbody radiation surface, the factory default emissivity is 0.95, and the emissivity can be adjusted between 0.96-0.99 if different emissivities are required.
- The screen has built-in multi-point independent temperature correction function.
- Automatic intelligent adjustment: Contains MH logic intelligent adjustment function, and advanced control algorithm for parameter self-tuning function.
- Program control function, the equipment has 10 groups of different program setting interfaces reserved, one-key operation, convenient and fast, meeting the user's requirements for multiple detection points and improving the user's detection efficiency.
- The communication interface can be customized RS232.
- The device has a memory function, which allows it to quickly restore to the previous setting state after restarting, without the need for additional settings to start.
Technical Indicators
Model
Temperature Range
Room temperature +5℃-300℃ / Room temperature +5℃-500℃
Operating Method
Touch Screen
Control Mode
PID Automatic Control
Temperature Resolution
0.1℃/0.01
℃
Radiating Surface Area
Φ150mm
Emissivity
≥0.95℃
Temperature Stability
Not greater than 0.15℃ and 0.15%t(bigger one in two,t is the raidiation source temperature
Temperature Uniformity
Assuming the radius of the target surface is r, all temperature points t within the half-price 2r/3 area must satisfy t-tc≤tc*0.5%+0.2℃, where tc is the temperature of the center point of the target surface (within the 2/3 center area)
Power Supply
220V AC 10A
Power
800W
Wire diameter
≥1.5 square
Volume (length × width × height)
Host: 215mm × 240mm × 290mm
Temperature controller: 300mm × 280mm × 130mm
Communication
RS232
Emissivity adjustable (adjustable between 0.92~0.99)
Weight
Host: 15.2Kg Temperature controller: 3.2Kg
Specifications
| Model | |
|---|---|
| Temperature Range | Room temperature +5℃-300℃ / Room temperature +5℃-500℃ |
| Operating Method | Touch Screen |
| Control Mode | PID Automatic Control |
| Temperature Resolution | 0.1℃/0.01 ℃ |
| Radiating Surface Area | Φ150mm |
| Emissivity | ≥0.95℃ |
| Temperature Stability | Not greater than 0.15℃ and 0.15%t(bigger one in two,t is the raidiation source temperature |
| Temperature Uniformity | Assuming the radius of the target surface is r, all temperature points t within the half-price 2r/3 area must satisfy t-tc≤tc*0.5%+0.2℃, where tc is the temperature of the center point of the target surface (within the 2/3 center area) |
| Power Supply | 220V AC 10A |
| Power | 800W |
| Wire diameter | ≥1.5 square |
| Volume (length × width × height) | Host: 215mm × 240mm × 290mm Temperature controller: 300mm × 280mm × 130mm |
| Communication | RS232 Emissivity adjustable (adjustable between 0.92~0.99) |
| Weight | Host: 15.2Kg Temperature controller: 3.2Kg |